Accuracy and precision of body temperature measurement with infrared tympanic and axiliary thermometers compared with standard oral mercury thermometer
Ahmad Khosravi *, Mohammad Bagher Sohrabi, Hossein Haratipoor, Poune Zolfaghari
Abstract:   (7174 Views)

Background: Monitoring patients’ temperatures is an important aspect of clinical nursing. Oral temperature readings have long been considered as a gold standard. However, oral temperature measurements may be contraindicated in some patients, that is cases with level of consciousness and seizure. Axiliary and tympanic temperature monitoring are often the next choices. We decided to evaluate these methods compared with standard oral temperature readings.

Materials and Methods: Subjects were a convenient sample of 81 adult patients over 17 years old in Shahrood Imam Hussein hospital. The study time course spanned a 1-mounth in December, 2004. Body temperature is measured in the evening using mercury oral and axiliary thermometers and “Omeron” tympanic thermometer.

Results: From 81 patients studied, there were 35(43.2%) males and 46(56.8%) females and their mean age was 46.0 ±19.2 (range 17-85). The mean for oral temperature was 36.9±0.56°C, axiliary: 36.5±0.53 °C, tympanic: 36.5±0.49° C. Oral thermometry identified a fever in 8 of 81 (9.9%) patients. The ROC analysis for determining a cut point with 100% sensitivity is used. For axiliary measurement at 36.95 °C, 82% specificity and for tympanic measurement at 36.45 °C, 55% specificity is recorded. Alpha coronbach for reliability of tympanic thermometer is equal to 0.98. The correlation between right to right and left to right ear temperature were 0.96 and 0.94, respectively. Conclusion: We conclude that axiliary thermometer with carefully use and suitable cut point is an effective instrument. With regard to specificity of tympanic thermometer for detection of fever, we do not recommend electronic tympanic measurement.

Keywords: Accuracy, Precision, Oral thermometer, Axiliary thermometer, Tympanic thermometer
Full-Text [PDF 324 kb]   (3175 Downloads)    
Type of Study: Research | Subject: General
Received: 2008/10/29
Add your comments about this article
Your username or Email:

CAPTCHA


XML   Persian Abstract   Print


Download citation:
BibTeX | RIS | EndNote | Medlars | ProCite | Reference Manager | RefWorks
Send citation to:

Khosravi A, Sohrabi M B, Haratipoor H, Zolfaghari P. Accuracy and precision of body temperature measurement with infrared tympanic and axiliary thermometers compared with standard oral mercury thermometer. Feyz. 2006; 9 (4) :43-49
URL: http://feyz.kaums.ac.ir/article-1-116-en.html


Volume 9, Issue 4 (Quarterly 2006) Back to browse issues page
مجله علمی پژوهشی فیض ::: دانشگاه علوم پزشکی کاشان KAUMS Journal ( FEYZ )
Persian site map - English site map - Created in 0.05 seconds with 32 queries by YEKTAWEB 3977